ge |
|
|
![]() |
||
|
The VT1 encapsulates the means of measuring vibration displacement, velocity, acceleration into a calculator format instrument incorporating rechargeable battery power source for total portability plus autoranging digital display of level over three decades. Signal source is a fixed sensitivity piezo-electric accelerometer a variety of transducers and accessories to facilitate VT1 operation are available separately, described under the P.V.A. Operation comprises simply applying the transducer to the test object, selecting the parameter required, and noting the reading, always presupposing that the data is within the VT1’s bandwidth/dynamic range window. IMPLEMENTATION APPLICATION NOTES Magnetic mountings, probes, degrade the frequency response. Transducers are directional, Qt ≡ g x cosθ, the axial error between the vibration vector and transducer axis, having implications in the application of probes. Autoranger update time and AC/DC converter averaging time, interacting with the dynamics of the physical signal, may produce large display perturbations ; resolvable perhaps by manual averaging or slowing the update/averaging parameters. Peak analogue signal level of 4.5V imposes a worst case, full scale display reading, crest factor Vpeak/Vrms of 2.25, equivalent to a 20% duty cycle rectangular wave signal. This level of crest factor introduces <1% additional conversion error.
|
|
|